日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
脳血流定量自動解析ソフト(AQCEL)の再構成におけるOS-EMパラメータの検討
浅尾 喜美枝竹田 ひとみ高木 昭浩河上 一公寺岡 悟見
著者情報
キーワード: ML-EM, OS-EM, AQCEL, iteration, subset
ジャーナル フリー

2008 年 64 巻 7 号 p. 822-831

詳細
抄録
The parameters of reconstruction iteration and subset of the ordered subset expectation maximization (OS-EM) method have a great impact on image quality. In a cerebral blood flow (CBF) examination, it is critical to produce consistent results when analyzing clinical data. We evaluated the number of iterations and subset of the OS-EM method to stabilize image quality using a normal case and an infarction case. In the maximum likelihood expectation maximization (ML-EM) method, it was confirmed that the convergence of an infarction case was somewhat delayed compared with a normal case. With the OS-EM method, we have obtained the same results. Based on the rCBF values, we postulate that the iteration value is more than 8 and the subset value is less than 10. Furthermore, multiplying the iteration by a subset value ranging from 50 to 60 helps in stabilizing the quality of CBF imaging.
著者関連情報
© 2008 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top