心臓MDCTにおける被ばく低減を目的としたハイヘリカルピッチ撮影の有用性
佐野 始也松谷 英幸近藤 武関根 貴子新井 雄大森田 ひとみ高瀬 真一
著者情報
ジャーナル フリー

2009 年 65 巻 7 号 p. 903-912

詳細
抄録

Helical pitch (HP) usually has been decided automatically by the software (Heart Navi) included in the MDCT machine (Aquilion 64) depending on gantry rotation speed (r) and heart rate (HR). To reduce radiation dose, 255 consecutive patients with low HR (≤60 bpm) and without arrhythmia underwent cardiac MDCT using high HP. We had already reported that the relationship among r, HP, and the maximum data acquisition time interval (Tmax) does not create the data deficit in arrhythmia. It was represented as Tmax= (69.88/HP-0.64) r; (equation 1). From equation 1, HP=69.88 r/ (Tmax+0.64 r); (equation 2) was derived. We measured the maximum R-R interval (R-Rmax) on ECG before MDCT acquisition, and R-Rmax×1.1 was calculated as Tmax in consideration of R-Rmax prolongation during MDCT acquisition. The HP of high HP acquisition was calculated from equation 2. In HR≤50 bpm, Heart Navi determined r: 0.35 sec/rot and HP: 9.8, and in 51 bpm≤HR≤66 bpm, r: 0.35 sec/rot and HP: 11.2. HP of the high HP (16.4±1.2) was significantly (p<0.0001) higher than that of Heart Navi HP (10.9±0.6). The scanning time (6.5±0.6 sec) of high HP was significantly (p<0.0001) shorter than that of Heart Navi (9.0±0.8 sec), and the dose length product of high HP (675±185 mGy⋅cm) was significantly (p<0.0001) lower than that of Heart Navi (923±252 mGy⋅cm). The high HP could produce fine images in 251/255 patients. In conclusion, the high HP acquisition is useful for reduction of radiation dose and scanning time.

著者関連情報
© 2009 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top