日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
原著
MPG印加軸数が拡散テンソル解析に及ぼす影響について
浜口 明巧小寺 秀一
著者情報
キーワード: MRI, diffusion, DTI, FA, ADC
ジャーナル フリー

2009 年 65 巻 7 号 p. 913-920

詳細
抄録
When diffusion tensor was analyzed, increasing motion probing gradient (MPG) directions and increasing signal-to-noise ratio (SNR) as result might influence diffusion tensor calculation. This study determined that mean diffusivity (MD) and fractional anisotropy (FA) are calculated by changing MPG-directions or SNR in the volunteer. There were no statistically significant differences in MD, which is calculated from all scanning parameters (p>0.05). FA of the caudate nucleus was similar as well. However, there were statistically significant differences between FA calculated from 31 MPG-directions and from 13 MPG-directions or less for the frontal lobe and caudate nucleus (p<0.05). If SNR of a trace image is 30 or more, FA wasn’t affected by SNR (p>0.05). To calculate FA, minimum MPG-directions that were not statistically different compared with 31 MPG-directions were 15 MPG-directions (p<0.05).
著者関連情報
© 2009 公益社団法人 日本放射線技術学会