日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
一般X線撮影時の照射条件設定システムの構築
小川 憲一
著者情報
ジャーナル フリー

2010 年 66 巻 8 号 p. 885-892

詳細
抄録

The purpose of this study was to determine the exposure conditions for general X-rays by calculation. The concept of the basic equation for the calculation is that the dose after transmitting an object is determined by entering the exposure conditions. For this purpose, we formulated a basic equation by mathematically representing all factors that influence imaging conditions. Using the basic equation, we succeeded in determining the optimal exposure condition for a patient’s body thickness by calculation. Even if another equipment, grid, added filter, etc., is used, optimal exposure conditions can be determined by adding correction factors determined by simple added measurements; this confirmed the versatility of the system. The calculation accuracy was verified to be within 15%. By using this system, we succeeded in optimizing exposure conditions for general X-rays.

著者関連情報
© 2010 公益社団法人 日本放射線技術学会
前の記事 次の記事