前立腺IMRTに用いるCBCTの線量評価と局所被ばく低減の対策
江崎 徹根本 幹央中根 義典盛満 寛乃増渕 二郎
著者情報
ジャーナル フリー

2011 年 67 巻 3 号 p. 183-192

詳細
抄録

Intensity modulated radiation therapy (IMRT) requires a precise set-up for which 3D matching using cone beam computed tomography (CBCT) is useful. Because images are acquired before every irradiation, investigating the exposure dose is essential. Such studies have been reported, but fewer reports have attempted to compare the dose with those in other modalities, as we did in this study. The exposure doses for imaging using 3 devices, including on board imager(OBI) CBCT, were measured using a tough water phantom, a self-made phantom, and a radiophotoluminescence glass dosimeter. The highest measured value was observed on OBI CBCT: the exposure dose was 4.3 times higher at some measurement sites than that from CT. Because this phenomenon is derived from the projection angle, a localized increase in the exposure dose may be prevented by changing the projection angle on each acquisition. Because the exposure dose for OBI CBCT was the highest and acquisition is performed before every irradiation, procedural modification is necessary, such as minimizing acquisition and the use of ultrasonography to confirm urine retention.