マルチディテクタCTにおけるハイブリッド型逐次近似画像再構成(iDose)アルゴリズムを用いた低線量撮像時の冠動脈石灰化の定量評価:フィルタ補正逆投影法との比較
村崎 裕生船間 芳憲羽手村 昌宏藤岡 知加子冨口 静二
著者情報
ジャーナル フリー

2011 年 67 巻 4 号 p. 360-366

詳細
抄録

To evaluate the usefulness of hybrid iterative reconstruction (iDose) for quantification of calcium content in the coronary artery on 64-detector computed tomography (CT), an anthropomorphic cardiac CT phantom containing cylinders with known calcium content was scanned at tube current-time products of 15, 20, 25, and 50 mAs using 64-detector CT. The images obtained at 15, 20, 25, and 50 mAs were reconstructed using filtered back projection (FBP), and those at 15, 20, and 25 mAs were also reconstructed using iDose. Then the volume and mass of the calcium content in the cylinders were calculated and compared with the true values. The Agatston score was also evaluated. The Agatston score and mass of calcium obtained at 50 mAs using FBP were 656.92 and 159.91 mg, respectively. In contrast, those obtained at 25 mAs using iDose were 641.91 and 159.05 mg, respectively. No significant differences were found in the calcium measurements obtained using FBP and iDose. In addition, the Agatston score and mass of calcium obtained at 15 mAs and 20 mAs using iDose were not significantly different from those obtained at 25 mAs with iDose. By using iDose, accurate quantification of calcium in the coronary artery can be achieved at 15 mAs using 64-detector CT. The radiation dose can be significantly reduced in coronary artery calcium scoring without impairing the detection and quantification of coronary calcification.