診療画像中に発生した黒点の原因となる放射性同位元素の付着部位に関する推定実験
西原 貞光林 裕晃花光 宏樹森 美智子
著者情報
ジャーナル フリー

2012 年 68 巻 10 号 p. 1307-1313

詳細
抄録

Caused by the accident of nuclear power plants in the Fukushima at 2011, many radioisotopes (RI) were diffused to the environment. As a result, X-ray detectors were stained with RIs and black spots appeared on the medical images. Using the RI of 134Cs and 137Cs, black spots which appeared on the photostimulable phosphor plate (X-ray detector) were reproduced experimentally. The aim of this study is the following two points; firstly, to clarify the relationship between long-time irradiations of RI and fading effect, and secondly, to clarify the positional relationship between the RI sources and the X-ray detector based on irradiation times of RI. For the latter experiment, the samples were made by spraying water (containing the RI) in order to reproduce small point sources. Then, the sources were placed on the photostimulable phosphor plate or on the cassette, and corresponding images with different irradiation times were taken. The black spots could be reproduced with the condition, in which sources were directly adhered to the photostimulable phosphor plate. We observed the black spots when sources were placed on the cassette for one week. Based on the result, we summarized that the RI which are directly adhered on the photostimulable phosphor plate may produce the black spots.