VMATに要求されるBackup JAWの位置精度と QA方法に関する検討
辰己 大作中田 良成家永 晃功四方田 あかね井上 誠市田 隆雄細野 雅子
著者情報
ジャーナル フリー

2012 年 68 巻 11 号 p. 1486-1491

詳細
抄録

The tolerance of the Backup diaphragm (Backup JAW) setting in Elekta linac was specified as 2 mm according to the AAPM TG-142 report. However, the tolerance and the quality assurance procedure for volumetric modulated arc therapy (VMAT) was not provided. This paper describes positional accuracy and quality assurance procedure of the Backup JAWs required for VMAT. It was found that a gap-width error of the Backup JAW by a sliding window test needed to be less than 1.5 mm for prostate VMAT delivery. It was also confirmed that the gap-widths had been maintained with an error of 0.2 mm during the past one year.