2種類の歯科用コーンビームCTの吸収線量と実効線量に関する検討
岡野 恒一松尾 綾江後藤 賢一横井 みどり蛭川 亜紀子奥村 信次小山 修司
著者情報
ジャーナル フリー

2012 年 68 巻 3 号 p. 216-225

詳細
抄録

Objective: This research compared the field of view with the tissue absorbed dose and effective doses using the two dental cone beam computed tomography (CBCT) scanners. Materials and Methods: Two CBCT devices, an Alphard VEGA and 3DX multi-image micro CT MCT-1, were used. Measurements were made using an Alderson RANDO phantom and thermoluminescence dosimeters (TLDs). The calculation of the effective dose was carried out according to ICRP Publication 60 and 103. Results: The effective doses for Alphard VEGA D mode, I mode, P mode, and C mode were 86, 238, 413, and 323 μSv, respectively. The effective doses using 3DX for the maxillary incisor, maxillary molar, mandibular incisor, mandibular molar, TMJ, and mandibular molar scout images were 27, 30, 48, 60, 14, and 1 μSv, respectively. Conclusions: Both Alphard VEGA and 3DX values revealed salivary gland and oral mucosa doses much higher than those required for other tissues. It is necessary to select a small mode suitable in order to realize optimization.