320-row Multi Detector Computed Tomographyにおけるノンヘリカルスキャンの物理特性
瓜倉 厚志中屋 良宏市川 勝弘川谷 圭亮川嶋 一平後藤 弘徳
著者情報
ジャーナル フリー

2012 年 68 巻 4 号 p. 432-442

詳細
抄録

Recently, clinical applications utilizing 320-row multi detector computed tomography (320MDCT) have increased, and the physical image properties of 320MDCT have been more concerned. We evaluated the spatial resolution in scan plane and z-direction, image noise and low-contrast sensitivity of non-helical mode (320NH), 640 slices mode by a double slice reconstruction technology (640DS), and 64-row helical mode (64HE) by using a 320MDCT. The spatial resolution in z-direction was evaluated by the section sensitivity profile (SSP) measurement with the micro coin phantom and the contrast transfer ratio (CTR) with the 0.5-mm comb phantom. The in-plane spatial resolution of 320NH was uniform over all the slice positions. The spatial resolution in z-direction decreased from the cathode side toward the anode side. The image noise of the anode side was higher than that of the cathode side. The contrast to noise ratio as index of the low contrast sensitivity was uniform over all the slice position. The CTR of 320NH fluctuated in the z-position, and the fluctuation was improved by 640DS except for the center of rotation.

著者関連情報