PET/CT検査におけるμ-mapのトランケーションエラーが体幹部のStandardized Uptake Valueに与える影響
甲谷 理温大西 英雄佐内 弘恭三村 浩朗柳元 真一
著者情報
ジャーナル フリー

2013 年 69 巻 2 号 p. 178-183

詳細
抄録

Purpose: The aim of this study was evaluate to impact of standardized uptake value (SUV) on the body trunk with truncation error of μ-map for CT attenuation correction (CTAC) in whole-body 2-deoxy-2-[18F] fluoro-D-glucose (18F-FDG)-positron emission tomography (PET)/CT with use of anthropomorphic phantom. Methods: We used body phantom (2.5 MBq/l) including simulated tumor targets (11.25 MBq/l) and arm phantom. The CT scan was used with a field of view (FOV) of 50 cm. The μ-maps were created by assuming a state of the arm protruding from the FOV (Pmap). A 3D-PET scan with an emission time of 20 min was performed. The PET images were then reconstructed with CTAC, and with and without scatter correction. We evaluated the relationship to Pmap size and the count of simulated tumors and background (B.G.) in PET images which reconstructed the use of each Pmap, respectively. Results: The count of simulated tumor (large) with scatter correction was decreased to 1.3% (Pmap: 15 mm) and 8.8% (Pmap: 35 mm). Then, the count severe reduction was 86.9% in Pmap of 65 mm. The same trend was shown by simulated tumor (middle, small) and B.G. The count of the simulated tumor (large) without scatter correction decreased to 1.3% (Pmap: 15 mm), 6.4% (Pmap: 35 mm) and 13.1% (Pmap: 65 mm). Conclusion: Truncation error by μ-map for CTAC in whole-body 18F-PET/CT caused a decrease of the SUV on the body trunk used for attenuation and scatter correction in the PET images.

著者関連情報
© 2013 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top