Hybrid O.R. systemにおけるステントグラフト内挿術の放射線被ばく
芳賀 喜裕千田 浩一加賀 勇治齋藤 和久荒井 剛鈴木 新一岩谷 佳美熊坂 江里子片岡 望佐藤 直人阿部 美津也
著者情報
ジャーナル フリー

2015 年 71 巻 12 号 p. 1215-1220

詳細
抄録

In recent years, aortic aneurysm treatment with stent graft grafting in the X-ray fluoroscopy is increasing. This is an endovascular therapy, because it is a treatment which includes the risk of radiation damage, having to deal with radiation damage, to know in advance is important. In this study, in order to grasp the trend of exposure stent graft implantation in a hybrid operating room (OR) system, focusing on clinical data (entrance skin dose and fluoroscopy time), was to count the total. In TEVAR and EVAR, fluoroscopy time became 13.40±7.27 minutes, 23.67±11.76 minutes, ESD became 0.87±0.41 mGy, 1.11±0.57 mGy. (fluoroscopy time of EVAR was 2.0 times than TEVAR. DAP of EVAR was 1.2 times than TEVAR.) When using the device, adapted lesions and usage are different. This means that care changes in exposure-related factors. In this study, exposure trends of the stent graft implantation was able to grasp. It can be a helpful way to reduce/optimize the radiation dose in a hybrid OR system.