偏重MPG三軸同時印加DWIの特性とADC値
山口 奈津美小倉 明夫林 則夫小山 大輔長壁 航己秦野 勇
著者情報
ジャーナル フリー

2015 年 71 巻 7 号 p. 595-604

詳細
抄録
The diffusion weighted images (DWIs) are used widely in clinical practice. Recently, the gradient overplus imaging was developed in addition to orthogonal technique for the duration method of motion-probing gradient pulses. The gradient overplus technique can reduce the load of a gradient strength, and enables setting of the shortest TE compared with the orthogonal gradients, and moreover it has the information on diffusion tensor. The apparent diffusion coefficient (ADC) values, signal-to-noise ratio (SNR), and geometric distortion were compared for the two duration methods of DWI using four different phantoms of coefficient of viscosity. In the gradient overplus, the time of smallest TE becamesmalle r than theorthogonal gradients according to theincre aseof theb value. As a result, SNR increased by shortening of TE, and thus the geometric distortion was improved. Moreover, the ADC value did not show any difference in the two duration methods of DWI. The gradient overplus technique is a more useful techniquein a clinical study than theorthogonal gradients becausethequality of imageis improved, and theADC valueis thesameas theorthogonal gradients.