Interventional Radiology 装置の線量率測定における留意点
吉田 秀義高橋 千春成田 信浩水沢 康彦関谷 勝大久保 真樹
著者情報
ジャーナル フリー

2016 年 72 巻 1 号 p. 63-72

詳細
抄録
In equipment used for interventional radiology (IVR), automatic exposure control (AEC) is incorporated to obtain the X-ray output suitable for the treatment of targeted lesions. For the AEC, users select a region as the signal sensing region (measuring field, MF) in the flat panel detector; MFs with various sizes and shapes were pre-defined and prepared in the system. The aim of this study was to examine the change of measured dose rate with the selection of MFs, the type of dosimeters (the ionization chamber dosimeter and the semiconductor dosimeter), and the dosimeter placement relative to the direction of X-ray tube (from cathode to anode). The IVR equipment was Allura Xper FD20/10 (Philips Medical Systems), and six kinds of built-in MFs were used. It was found that dose rate measured by the ionization chamber dosimeter showed a variation of -2 mGy/min with the MFs and the ionization chamber dosimeter placement. The dose rate measured by the semiconductor dosimeter showed more variation than the ionization chamber dosimeter. The change of dose rate with the dosimeter placement would be caused by the MF overlapping the dosimeter which would affect the AEC (the X-ray output). Also, the change of dose rate with the dosimeter placement was considered to be related to the heel effect of the X-ray beam. When performing dose rate measurements, we should notice that the selection of MFs, the type of dosimeters, and the dosimeter placement would affect the measured values.
著者関連情報
© 2016 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top