日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
造影剤濃度と注入プロトコルの違いが前立腺dynamic contrast enhanced MRI に与える影響
牧野 佑也秋元 聰星野 広史
著者情報
ジャーナル フリー

2018 年 74 巻 7 号 p. 685-691

詳細
抄録

The purpose of this study was to evaluate the enhancement profile of 1.0M gadobutrol (high concentration: HC) in comparison to 0.5M gadopentetate dimeglumine and gadoterate meglumine (low concentration: LC) in dynamic contrast enhanced magnetic resonance imaging (DCE-MRI) of prostate. In total, 48 patients who were diagnosed with prostate cancer by radiologist were included. Each patient was examined after intravenous injection of 0.1 mmol/kg body weight contrast agent with flow rate of 1.5 (HC) or 3.0 mL/s (LC). Circular regions of interest were placed at prostate cancer (PCa) and normal peripheral zone (normal PZ) in DCE-MRI. The enhancement curves were calculated as a relative enhancement. Statistical analysis was performed by Mann-Whitney U test (p<0.05 were considered significant). As a result, the enhancement at first phase of HC was significantly lower compared with LC in PCa (HC, 0.47; LC, 0.85; p=0.029), and in normal PZ (HC, 0.12; LC, 0.22; p=0.033). The enhancement of HC in PCa was significantly higher compared with LC at late phase. Although not significant, a similar tendency was observed in normal PZ. The present study suggested that the enhancement profile with HC was higher at late phase but the rise of the enhancement curve with HC tended to be delayed.

著者関連情報
© 2018 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback