脳神経血管および心臓電気生理手技のインターベンションに携わる医師の水晶体線量評価
加藤 守千田 浩一石田 嵩人佐々木 文昭豊嶋 英仁大阪 肇木下 俊文
著者情報
ジャーナル フリー

2020 年 76 巻 1 号 p. 26-33

詳細
抄録

Purpose: The International Commission on Radiological Protection recommended that interventional radiologies (IRs) have high radiation doses and that staff may also be exposed to high doses. In the present study, we measured the radiation exposure dose [3 mm dose equivalent, Hp (3) ] in the eye using an appropriate dosimeter placed next to the physician’ s eye during neurovascular intervention procedure (Neuro-IR) and interventional cardiac electrophysiology procedure (EP-IR). Method: Physicians wore a direct eye dosemeter just lateral to the left eye and an additional direct eye dosemeter outside the radiation protective glasses close to their left eye. Additionally, a neck badge [0.07 mm dose equivalent, Hp (0.07) ] was worn outside the protective apron to the left of the neck, to compare the direct eye dosimeter estimated doses. The occupational eye lens dose was evaluated over a period of 6-month. Results: The maximum Hp (3) of the Neuro-IR physician was estimated 5.1 mSv without the radiation protective glasses and 1.6 mSv with the radiation protective glasses. On the other hand, the maximum Hp (3) of the EP-IR physician was estimated 29 mSv without the radiation protective glasses and 15 mSv with the radiation protective glasses. Conclusion: Physicians eye lens dose [Hp (3) ] tended to be overestimated by the neck badge measurements [Hp (0.07)]. A correct evaluation of the lens dose [Hp (3) ] using the direct eye dosimeter is recommended. Although we found a positive correlation between Hp (0.07) and Hp (3), the value of R2 in the regression equation is low, we recommended that the eye lens dose estimated carefully from Hp (0.07).

著者関連情報
© 2020 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top