Dual energy CT を用いた仮想単色X 線画像におけるビームハードニングの改善効果の評価:電子密度ファントムを用いた検証
平入 哲也市川 勝弘瓜倉 厚志杉山 正則阿瀬川 敏松波 環
著者情報
ジャーナル フリー

2020 年 76 巻 2 号 p. 168-176

詳細
抄録

Purpose: A virtual monochromatic image (VMI) is acquired from two different types of polychromatic energy X-rays, not a monochromatic X-ray. The effective energy of monochromatic X-ray does not vary in passing through the patient’s body. On the other hand, beam hardening effects are seen in images because of the change of polychromatic X-ray energy. The purpose of the present study was to evaluate the beam hardening improvement effect of VMI using a phantom with a bone mimicking ring. Method: We used a water equivalent electron density phantom with a hole in the center for inserting various measurement materials (i.e. fat, two types of bone with differing densities, contrast medium, blood, and water). Then, the CT numbers of each measurement materials were obtained from single energy CT (SECT) images and VMIs, respectively. Also, an additional bone-mimetic ring was used to obtain the CT numbers for evaluation of beam hardening effect. The CT number change rates were calculated from the obtained CT numbers with and without beam hardening effect. Result: The rate of CT number, change of VMI was significantly lower than that of SECT for all measured materials. Conclusion: In this study, VMI minimized changes in CT numbers due to the beam hardening effect and showed a higher beam hardening reduction effect.