CsI-flat panel detector を用いる胸部撮影における90 kV,0.15 mm Cu フィルタ線質の評価
小田 敍弘黒川 善之上原 周三佐々木 範男山縣 久仁子山崎 慎矢田吹 晃一横田 信弥亀山 伸吉井原 宏彰中野 努
著者情報
ジャーナル フリー

2020 年 76 巻 5 号 p. 463-473

詳細
抄録

Purpose: To compare the visibility of anatomic structure in chest radiography acquired with different beam quality (120 kV beam and 90 kV beam with 0.15 mmCu) using CsI-flat panel detector. Method: Pair image obtained by different beam quality of 100 person's chest radiographies which were taken periodical health examination were compared with the visibility of normal structures (pulmonary vessels) and abnormal opacities by two pulmonologists and four radiological technologists. Moreover, the spectrum of the two beam quality were calculated using Monte Carlo simulation. Result: Dominant observers gave high score significantly (p<0.01) to the 90 kV beam's image in spite of 20% less dose. Monte Carlo simulation showed that 90 kV beam with 0.15 mmCu were much absorbed primary photon than 120 kV beam to CsI detector, and less absorbed secondary photon. Conclusion: The visibility of anatomic structure and abnormal opacities in FPD chest radiography was improved by using the 90 kV beam with 0.15 mmCu than traditional 120 kV beam's chest radiography.

著者関連情報
© 2020 公益社団法人 日本放射線技術学会