移動型整形外科用イメージング装置を用いた 腰椎後方固定術時における室内散乱線量分布の測定
福永 正明松原 孝祐竹井 泰孝光井 英樹亀井山 弘晃田中 孝尚市川 翔太
著者情報
ジャーナル フリー

2020 年 76 巻 6 号 p. 572-578

詳細
抄録

The purpose of this study was to measure the scatter radiation intensity during transforaminal lumbar interbody fusion using a mobile C-arm system (Arcadis Orbic 3D; Siemens) and minimize radiation exposure. Dosimetry was performed with anterior-posterior and lateral continuous fluoroscopy, and cone beam computed tomography (CT). A scaffold tower (L: 300 cm×W: 200 cm×H: 150 cm) was built with radiation-resistant paper cylinders at intervals of 50 cm and plastic joints over the bed, and 100 optically stimulated luminescence dosimeters (nanoDot; Nagase Landauer) were placed on each joint. A human torso phantom from head to pelvis (Kyoto Kagaku) was positioned on the bed in a prone position. The scatter radiation dose in a lateral view was highest on the X-ray tube side at the height of 100 cm (170.5 μGy/min). The scatter radiation dose increased significantly on the X-ray tube side during lateral continuous fluoroscopy. Continuous change of surgeons' standing positions is important to minimize radiation exposure received by a specific surgeon.