2020 年 76 巻 9 号 p. 911-917
Although the test bolus tracking method is available as a predicting method of scan timing in the coronary computed tomography (CT) angiography, it is known that there is a problem of scan timing due to the use of a part of test bolus method. The diluted test bolus method was adopted for test bolus, and as a result of using in combination with the test bolus tracking method, it showed a higher contrast enhancement compared with the test bolus tracking method; a stable contrast enhancement with less variation in CT number was obtained. The CT number at the peak in the test scan and the CT number of the main scan showed a high correlation. The contrast injection technique using the diluted test bolus method and the test bolus tracking method is a useful method in the coronary CT angiography. We named this contrast injection technique diluted test bolus tracking method.