生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
抄録
単眼内視鏡による立体形状計測の高精度化
秋元 賢仁, 中島 義和, 谷垣 信行, 竹下 隆章, 金 ミン奎
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S203_03

詳細
抄録
Endoscopy measuring three-dimensional (3D) shape of organ inner wall is addressed. It can provide organ surface shape and texture simultaneously and improve diagnosis capability such as to find small gastric submucosal tumors. Some endoscpic systems based on stereo measurement or laser scanning have been proposed for this purpose, however, they have the drawbacks of corresponding unstability and occulusion between stereo images. In this paper, we demonstrate 3D shape measurement with single lens endoscopy by a new shape-from-focus (SFF) method that compensates error caused by blur spatial non-uniformity. Our system improved organ shape measurement from 1.38 mm to 0.53 mm RMS error.