Journal of the Vacuum Society of Japan
Online ISSN : 1882-4749
Print ISSN : 1882-2398
ISSN-L : 1882-2398
解説
帯電水滴衝撃イオン化/二次イオン質量分析法
平岡 賢三
著者情報
ジャーナル フリー

2009 年 52 巻 4 号 p. 237-241

詳細
抄録
  A new type of cluster SIMS, named as electrospray droplet impact (EDI), has been developed in our laboratory. It was found that peptides deposited on the stainless steel substrate were ionized/desorbed without the accumulation of radiation products. The organic samples with film thickness thinner than 10 monolayers are desorbed/ionized with little damage underneath the surface. In general, rather strong negative ions as well as positive ions are generated. The mechanism for the ionization/desorption in EDI is much less complicated than those for MALDI and SIMS due to the fact that only very thin sample layers take part in the shock-wave excited selvedge and higher-order side reactions are largely suppressed.
著者関連情報
© 2009 一般社団法人日本真空学会
前の記事 次の記事
feedback
Top