201TlCl 心筋 SPECT 像の検討 : 180 度と 360 度データ収集法の濃度均一性の比較
岡田 淳徳中條 政敬紫垣 誠哉堀江 有記石原 幸生村山 恵洋穂満 信行下野 哲勇
著者情報
ジャーナル フリー

1993 年 49 巻 9 号 p. 1601-1608

詳細
抄録

A comparative 360°/180°data aquisition study with a myocardial phantom was performed to detect a suitable data sampling technique for myocardial emission computed tomography using thallium-201. The homogeneous distribution of radioative is more frequently shown in the 180°data sampling technique than 360°data sampling techniqe for all the experimental studies, especially in the seputum and posterior/inferior areas than in other myocrdial areas. It was concluded that the 360°data sampling technique is superior to the 180°technique in evaluating myocardial perfusion imaging using thalllim-201.