X線シネ撮影装置の自動露出機構の被写体厚特性と応答特性
荒居 広明
著者情報
ジャーナル フリー

1997 年 53 巻 11 号 p. 1673-1682

詳細
抄録
X-ray characteristics including subject thickness(copper plate), tube voltage, tube current and irradiation time were measured at stability, with an automatic dose control x-ray generator for cineangiography. Regardless of subject thickness, it is possible that the energy input to the x-ray tube in one frame may be decreased. The automatic control response was measured after rapid fluctuation in subject thickness. Two inverter-type x-ray generators with different automatic control units were studied. The older control unit changes exposure by tube voltage and tube current, while the newer one changes exposure dose by tube voltage, tube current and irradiation time. The maximum rate of change in tube voltage is greater with the newer control unit. In addition, the actual tube current response of the newer control unit in increasing nominal value is faster than the older one. In the new control unit, for each pulse, irradiation is cut off by means of a signal that the exposure has reached the proper value. Thus given the same differential in subject thickness, the newer control unit resumed stability faster than the older one.