矩形波チャートを用いたディジタル系のMTF測定の検討
畑川 政勝井上 誠小川 隆由苅屋 公明
著者情報
ジャーナル フリー

1997 年 53 巻 11 号 p. 1683-1690

詳細
抄録
The presampling modulation transfer function(MTF)and the final MTF have been used in evaluating the resolution of digital imaging systems. These MTFs are generally measured by the slit method. On the other hand, in the analog system, the square wave response function(SWRF)method is widely used to measure MTF. In this study, we examined the use of the SWRF method with a digital imaging system. Results showed that output modulations of square waves were varied when data were sampled sparsely. However, the presampling MTF and the final MTF could be measured from the maximum and minimum values of output modulation. The measured final MTFs obtained from the slit method and the SWRF method were in good agreement with one another.
著者関連情報