漏洩X線の実効エネルギー計測に関する研究
藤本 信久大釜 昇西谷 源展山田 勝彦
著者情報
ジャーナル フリー

1997 年 53 巻 3 号 p. 373-379

詳細
抄録
It is important to correct for the energy response of the X-ray detector in measuring X-ray leakage from the protective barrier of the examination room. However, measurement of effective energy of leakage X-rays is very difficult in the clinical setting. In this paper, we deiscuss a method for estimating the effective energy of both primary and scattered X-rays. This method is based on the measurement of leakage X-ray spectra using a semiconductor-fitted X-ray detector and analyzing the relationship between X-ray tube voltage and effective energy. In this study, we found that the effective energy of leakage X-rays was not dependent on the thickness of shielding materials like concrete and lead. With the concrete barrier, it was possible to express the relationship between X-ray tube voltage and the effective energy of leakage X-rays with a simple approximate equation. With the lead-lined barrier, the absorption of the K-edge of lead needs to be considered in estimating the effective energy of leakage X-rays at relatively high tube voltages.