The MM50 is a racetrack microtron capable of taking out photon beams and electron beams with energies of up to 50 MeV. It flattens the beam by the beam-scanning method, while the microtron MM22 utilizes a flattening filter. The head-scatter factors (hereafter called S_h) , which are important for evaluating the output of the photon beam of the MM50 and MM22, were measured using a mini-phantom and build-up cap. Sh measured with the build-up cap showed the influence of contaminated electrons, whereas S_h measured with the mini-phantom showed less influence, even for 50 MV photon beams. Compared with the MM22, the MM50 showed less change in S_h according to field size and energy. The reason for this seemed to be that the MM50 has a smaller extra-focal region than other accelerators equipped with flattening filters and, therefore, can essentially be considered a point source by using the beam-scanning method without a flattening filter. This study demonstrated that photons scattered by the flattening filter used for beam flattening in typical medical accelerators mainly contribute to S_h.