MRI検査時におけるインプラントの生体への影響 : 金属球を用いた基礎的実験
村中 博幸中村 修笛吹 修治上田 善武森川 薫
著者情報
ジャーナル フリー

2005 年 61 巻 7 号 p. 1014-1020

詳細
抄録

It is increasingly the case that patients who have implants feel pain during high-field MRI examinations. A probable reason for the pain is the generation by irradiation of RF pulses and changing of the magnetic field gradient. As a fundamental study on the effect of implants on the human body under MRI procedures, temperature measurements were obtained from metal balls incorporated into gel-filled phantoms by using two kinds of measuring instruments, a copper-constantan thermocouple and a fluorescence fiber thermometer. At first we pursued a correlation between a copper-constantan thermocouple (absolute measurement) and fluoroptic thermometer and confirmed the precision and stability of the fluoroptic thermometer under MRI procedures. When a stainless steel ball with or without a loop antenna was used, only in the former case did the temperature rise during RF pulse irradiation. There was no significant difference between the magnetic field gradient ON and OFF. Furthermore, differences in metal (steel, aluminum, brass, stainless steel, copper) and size (5, 10, 20mmΦ) were affected according to the increase of temperature. In conclusion, both RF pulse irradiation and a loop antenna are necessary for heat generation on the surface of metals.