サブトラクション処理を用いたヘリカルアーチファクトの新しい評価法
西丸 英治宇都宮 律子西江 亨文二野宮 祐司友重 之宏
著者情報
ジャーナル フリー

2005 年 61 巻 7 号 p. 1021-1026

詳細
抄録

This paper proposes a new quantitative method of appraising the helical artifact generated by computed tomography (CT) images by helical scan. The quantities-appraisal method of the helical artifact uses as a sample image, the binarization image, which performs the threshold operation. Therefore, results may differ among observers and are lacking in objectivity. We devised a procedure (subtraction technique) that uses the subtraction process as a method of appraising the helical artifact that does not a perform a threshold operation, and evaluated this procedure. As compared with an area technique, a subtraction technique can divide a helical artifact into a plus CT value (Over Level) and a minus CT value (Under Level) and can evaluate them. This provides excellent repeatability and objectivity. Moreover, since CT value data are maintained, the quality of an artifact can also be evaluated, making this a very useful procedure. By using the subtraction technique, new information will be found by evaluating helical artifacts.

著者関連情報
© 2005 公益社団法人 日本放射線技術学会