ノート
3テスラ両側乳腺ダイナミックにおける脂肪抑制方法の比較:Three-point DIXONの有用性について
水戸 寿々子石坂 欣也中西 光広杉森 博行濱口 裕行仲 知保
著者情報
ジャーナル フリー

2011 年 67 巻 6 号 p. 654-660

詳細
抄録

The purposes of this study were to determine optimum flip angles (FAs) and to compare the effectiveness of fat suppression and signal homogeneity among three techniques, spectral attenuated with inversion recovery (SPAIR), principle of selective excitation technique (PROSET), and three-point DIXON technique (DIXON), of the bilateral breast dynamic sequence acquired using the optimum FA at 3.0 T. Using a homemade phantom that represented a tumor, fat, and a mammary gland, the optimum FAs were determined from the change of fat signal intensity, signal-to-noise ratio (SNR) of the mammary gland, and contrast ratio (CR) between the tumor and mammary gland. The effectiveness of fat suppression and signal homogeneity were compared in ten breast cancer cases, using the CR between fat and pectoralis muscle signal intensities and the standard deviation (SD) of fat signal intensity, respectively. The optimum FAs for SPAIR, PROSET, and DIXON were 10, 20, and 20 degrees, respectively. The mean CR between fat and pectoralis muscle signal intensities achieved using SPAIR, PROSET, and DIXON were 0.19, 0.30 and 0.40, respectively, and the mean SDs of the fat signal intensities were 90.2, 103.1, and 30.5, respectively. The DIXON technique provided better fat suppression and signal homogeneity than the other two techniques. The results of this study suggest the possible application of the DIXON technique in combination with the optimum FA setting as an effective fat suppression technique for the bilateral breast dynamic sequence at 3.0 T.

著者関連情報
© 2011 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top