日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
心電同期拡散MRIによる脳のADC測定時の最適心時相
菅 博人宮地 利明間瀬 光人笠井 治昌原 眞咲芝本 雄太川野 誠大野 直樹北中 章博依光 里沙南 貴司
著者情報
ジャーナル フリー

2011 年 67 巻 6 号 p. 661-665

詳細
抄録

Apparent diffusion coefficients (ADCs) are sometimes overestimated when they are measured in the brain near the basal ganglia because water molecules in brain tissues fluctuate with blood volume loading in the cranium. We determined detailed ADC changes during the cardiac cycle to evaluate the appropriate cardiac phases for accurate measurement of ADC values. Using 1.5 T MRI, we performed ECG-triggered single-shot EPI to obtain ADC maps in each cardiac phase using techniques minimizing the bulk motion effect. The coefficient of variation (CV) of the ADC values during the cardiac cycle was over 50% near the basal ganglia. Moreover, the cardiac phase of the peak ADC value during the cardiac cycle was from 10 to 40% cardiac phases that follow systole. However, the CV of the ADC values of whole cardiac phases was higher than those with phases over 50% of the cardiac cycle near the basal ganglia because the effect of water fluctuation was almost eliminated. Therefore, accurate measurement of ADC values should be obtained from ADC maps of phases over 50% of the cardiac cycle.

著者関連情報
© 2011 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top