日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
原著
GM計数管を用いたピンホールカメラの開発に向けた基礎研究
竹上 和希林 裕晃福田 郁磨小西 有貴花光 宏樹橋詰 拓弥村上 淳三谷 聡
著者情報
ジャーナル フリー

2013 年 69 巻 9 号 p. 944-951

詳細
抄録

To visualize the scattered X-ray distribution in the diagnosis domain, we examined whether a Geiger Mueller (GM) counter could be applied to a pinhole camera as an X-ray detector. The GM counter detects radiation at certain detection points. To obtain two-dimensional images using the GM counter, the detector needs to be moved two-dimensionally. We constructed an apparatus using industrial actuators to move the detector. To investigate the usability of the developed apparatus, the scattered X-rays from the phantom were measured using the GM counter. The images obtained were then compared with those measured using the phosphor plate. Our results demonstrated that the GM counter can detect low count-rate radiation, but further research will be needed to obtain clear two-dimensional images. In this paper, we propose that the GM counter can be used as a complementary detector to a phosphor plate.

著者関連情報
© 2013 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top