123I-MIBG心筋シンチグラフィにおける心縦隔比と洗い出し率の解析方法と再現性の検討
長谷川 大輔大西 英雄松友 紀和甲谷 理温
著者情報
ジャーナル フリー

2014 年 70 巻 12 号 p. 1420-1428

詳細
抄録

Purpose: The aim of this study was to investigate an appropriate analysis method and multicenter reproducibility of heart-to-mediastinum ratio (H/M) and washout rate (WR) in iodine-123-labeled metaiodobenzylguanidine (123I-MIBG) myocardial scintigraphy with a phantom. Methods: We evaluated the optimal region of interest (ROI) setting method about the mediastinum and heart by varying the position and shape of the ROI. The mathematical method was changed to a combination of decay time correction (DTC) and background correction (BC). We evaluated the reproducibility of the H/M and WR between institutions. Result: H/M decreased to 23.49% and WR increased to 20.68% by changing the mediastinum ROI position from upper to lower. H/M increased to 26.03% by changing the heart ROI position from base to apex. H/M decreased to 38.36% with BC, and WR was reduced up to 48.51% with DTC. Reproducibility of the H/M and WR between institutions was improved by performing optimization of the ROI setting and unification of the mathematical method. Discussion: The position of the mediastinum ROI should be set on the upper mediastinum. The position of the heart ROI should be set on the apex of the heart. WR should be calculated with DTC and BC. Our results suggest that the reproducibility of the H/M and WR between institutions was improved by performing optimization of the ROI setting and unification of the mathematical method.