コバルト60 定位手術的照射装置における線源位置変位が線量および線量分布に及ぼす影響
中澤 寿人内山 幸男
著者情報
ジャーナル フリー

2015 年 71 巻 3 号 p. 177-185

詳細
抄録

We assessed the impact of source positional discrepancy on dose and dose distributions in Gamma Knife (GK) Perfexion (PFX) stereotactic radiosurgery. A spherical phantom dedicated in GK machine was used and irradiated by 2 Gy in each position moved at an interval of 0.1 mm from its original position using three types of collimators (4, 8, 16 mm) to evaluate the changes of dose. In addition, to obtain the dose distributions, radiochromic film was inserted in the phantom and irradiated by 6 Gy in each position moved at an interval of 1 mm from its original position using three types of collimators. A distance-to-agreement analysis (DTA) was performed to compare isodose lines from 10% to 90% of dose distributions between the original and deviated position. As a result, when the source moved toward the discrepancy from the center of the collimator, the dose and dose distributions discrepancies increased according to the degree of discrepancy. Especially in 4-mm collimator, 0.5 mm discrepancy caused dose reduction of 5%. On the other hand, 0.5 mm discrepancy showed merely dose differences less than 0.5% in 8 mm and 16 mm collimators. Regarding dose distributions, 1 mm discrepancy in all collimators showed little changes in DTA within 1 mm on average.