We compared the uniformity of fat-suppression and image quality using three-dimensional fat-suppressed T1-weighted gradient-echo sequences that are liver acquisition with volume acceleration (LAVA) and Turbo-LAVA at 3.0T-MRI. The subjects were seven patients with liver disease (mean age, 66.7±8.2 years). The axial slices of two LAVA sequences were used for the comparison of the uniformity of fat-suppression and image quality at a region-of-interest (ROI) of the liver dome, the porta, and the renal hilum. To yield a quantitative measurement of the uniformity of fat suppression, the percentage standard deviation (%SD) was calculated by comparing two sequences. For image signal to noise ratio (SNR), the contrast between the liver and fat (Cliver-fat), and the liver and muscle (Cliver-muscle), the other ROIs were placed in the superficial fat, liver, spleen, pancreas, and muscle. The %SD in Turbo-LAVA (28.1±16.8%) was lower than that in LAVA (41.5±13.4%). The SNRs in Turbo-LAVA (17.8±4.1 [liver], 12.5±3.0 [pancreas], 14.7±1.6 [spleen], 8.2±3.5 [fat]) were lower than those in LAVA (20.9±6.1 [liver], 16.8±4.1 [pancreas], 17.4±2.4 [spleen], 12.0±4.5 [fat]). While, the Cliver-fat in the Turbo-LAVA (0.72±0.06) was significantly higher than that in LAVA (0.59±0.07). Turbo-LAVA sequence offers superior and more homogenous fat-suppression in comparison to LAVA sequence.