電子ポータル画像装置を用いた前立腺癌に対する強度変調回転治療中の線量評価
今江 禄一竹中 重治渡邉 雄一岡野 由香里根津 誠三枝 茂輝竹内 幸浩矢野 敬一芳賀 昭弘白石 憲史郎山下 英臣中川 恵一
著者情報
ジャーナル フリー

2016 年 72 巻 11 号 p. 1128-1136

詳細
抄録

Purpose: Volumetric modulated arc therapy (VMAT) is capable of acquiring projection images using electronic portal imaging device (EPID). Commercial EPID-based dosimetry software, dosimetry check (DC), allows in vivo dosimetry using projection images. The purpose of this study was to evaluate in vivo dosimetry for prostate cancer using VMAT. Method: VMAT plans were generated for eight patients with prostate cancer using treatment planning system (TPS), and patient quality assurances (QAs) were carried out with phantom. We analyzed five plans as phantom study and five plans as patient study. Projection images were acquired during VMAT delivery. DC converted acquired images into fluence images and used a pencil beam algorithm to calculate dose distributions delivered on the CT images of the phantom and the patients. We evaluated isocenter point doses and gamma analysis in both studies and dose indexes of planning target volume (PTV), bladder and rectum in patient study. Results and discussion: Dose differences at the isocenter were less than a criterion in both studies. Pass rates of the gamma analysis were less than a criterion by two plans in the phantom study. Dose indexes of reconstructed distribution were lower than original plans and standard deviations of PTV in reconstructed distribution were larger than original plans. The errors were caused by some issues, such as the commissioning of DC, variations in patient anatomy, and patient positioning. Conclusion: The method was feasible to non-invasively perform in vivo dose evaluation for prostate cancer using VMAT.

著者関連情報
© 2016 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top