ビスマスシートを用いたcomputed tomography fluoroscopyにおける術者被ばくと画質の評価
滝口 京佑瓜倉 厚志吉田 司廣澤 賢一伊東 孝宏中屋 良宏
著者情報
ジャーナル フリー

2016 年 72 巻 11 号 p. 1137-1143

詳細
抄録

Objective: The purpose of this study was to assess the dose reduction and the image quality using bismuth sheets during the computed tomography fluoroscopy (CTF). Materials and Methods: The bismuth sheets of 1-mm thick were put on the upper mylar ring to reduce the frontal X-ray. The dose rates of an operator were measured using a torso phantom in the patient position during the CTF. The torso phantom was set on the gantry rotation center (center) and the lower position from the center (off-center). The image quality of the CTF image was assessed using an original phantom that mimics the normal liver parenchyma and the low attenuation lesions. The image contrast and contrast-to-noise ratio (CNR) were compared with and without the bismuth sheets. Results: The bismuth sheets reduced the dose rate of the operator, regardless of whether the torso phantom was set at the center or the off-center. The reduction rate of exposure at the center and the off-center were 42.3% and 34.5%, respectively. There were no significant differences in the image contrast and the CNR, although the bismuth sheets increased the CT values of the liver parenchyma and the low attenuation lesions. Conclusion: The bismuth sheets were effective for the reduction of exposure to the operator without degrading the image quality of CTF images.