Iryou kikigaku (The Japanese journal of medical instrumentation)
Online ISSN : 1884-054X
Print ISSN : 1882-4978
ISSN-L : 1882-4978
Volume 94, Issue 5
Displaying 1-11 of 11 articles from this issue
Original Contribution
  • Masatoshi Koyama
    2024Volume 94Issue 5 Pages 495-499
    Published: 2024
    Released on J-STAGE: December 11, 2024
    JOURNAL FREE ACCESS

    Background: For NPPV machines that add oxygen, the administered oxygen concentration is unknown and the oxygen concentration may differ from the one supplied. Method: Using 3 models, we experimentally verified the oxygen concentration depending on the model, ventilation rate, and oxygen addition site by varying the supplied oxygen flow rate of 2 L/min, 6 L/min, and 10 L/min. Results: No differences were observed in oxygen concentration depending on model or ventilation volume. There were differences in oxygen concentration depending on the location of oxygen addition. Conclusion: Even with the same oxygen administration flow rate, the oxygen concentration differs depending on the oxygenation site.

    Download PDF (795K)
Short Report
Review / Proposals for better practical training in clinical engineering