Objective : This study aimed to clarify the influence of the thickness of the implant collar part and cantilever on the maxi mum bending load, deformation, and strain of the collar.
Materials and Methods : The superstructure was prepared by simulating the mesiodistal width (12 mm) of the lower first molar. To apply a load, 3.25 mm was added to both ends, resulting in a full length of 18.5 mm (using JIS SK4 steel). The thickness of the collar part was changed to 0.7 (T07), 0.8 (T08), or 0.9 (T09) mm (using JIS 4 type titanium). The angle of inclination of the implant was set to 30°, and the maximum bending load, deformation, and strain were measured. The inner region was observed using CT.
Results and Discussion : The maximum bending load was 387.3, 402.9, and 414.5 N for T07, T08, and T09, respectively ; thus, the maximum bending load increased as the implant collar became thicker (p<0.05). The measured deformation was 1.86, 1.82, and 1.52 mm for T07, T08, and T09, respectively ; thus, T09 had the least deformation (p<0.05). Strain of 0.1% was caused by a 150 N load in both T07 and T08, and by a 200 N load in T09. Strain of 0.3% was caused by a 350 N load in both T07 and T08. The strain caused by a 400 N load in T09 was less than 0.3% (p<0.05). Computed tomography (CT) imaging further revealed that a portion of the abutment of each implant had deformed and invaded into the chip evacuation hole. There was a gap between the tip of the collar and abutment that fit the collar, and the gap was the greatest for T07.
Conclusion : Using implants with a cantilever, we demonstrated that the thickness of the collar affected the degree of strain. This indicated that the biting force at the site of an implant should be carefully considered.
View full abstract