The cyclotron framed with many equipments has been inspected already and vacuum leakage test was tried for the first time with both the propane and the helium gas detectors for a month.
After seventeen leaks detected were treated eight times, the final chamber pressure became 1.1×10
-5mmHg and the total quantity of liberated gas was 1.0×10
-3mmHg·1· sec
-1 by the air scale of the ionization gauge.
This amount of librated gas per cm
2 per sec, was less than that of gas evolved from copper parts only, about which measurement had been done before assembling and the spectra fron Geissler discharge tube shew a nitrogen band only.
Analysing this work, we found followings;
a) the total quantity of leaked air was expressed by a negative exponential curve against the times of repair, b) the detected leakage being compared with the standard capillary leak of glass, a leakage value estimated with the propane gassingle ionization gauge system was about one third of the one estimated with the helium gas-spectrometer system.