A new flicker test apparatus has been developed, by which measurements of critical flicker frequency (CFF) can be repeated stably in a short duration, and the method of measurement using this apparatus was investigated. A microcomputer used to control the flicker frequency, the treatment of the response switch and the saving data in this apparatus. Controlling flicker frequency up and down repeatedly, the two measured curves were obtained at one time, the one was the CFF curve by the ascending method (fu-CFF), and the other was by the descending method (fl-CFF).
The curves showed patterns whose initial values were high, followed by gradual decrease over a short period of measurement. The fl-CFF of females were higher than that of males, the drop of CFF during the test of the females were smaller than that of males. According to the experiment loading the task by the Kraepelin test, the changes were recognized in the initial CFF which corresponded to the mean value by the conventional method, in the fluctuation of CFF, and in the difference between fu-CFF and fl-CFF(hysteresis). The remarkable change of the drop of CFF with no change of the initial CFF was found in one case.
The changes of hysteresis, the fluctuation of CFF and the drop of CFF with the task were obtained, which could not get by the conventional method. Therefore, the method of measurement using the Continuous Flicker Test Apparatus considered to be useful as a new functional test for evaluating human stress resulting from work.
抄録全体を表示