Purpose: To optimize patient radiation dose in Spiral CT scan of dento-maxillo-facial region by measuring the absorbed dose in the phantom and to evaluate reliability of dose estimation methods using CTDI (CT Dose Index, FDA, USA).
Method: Spiral CT scanning with “pitchs” (ratio of table speed to slice thickness per rotation) more than 1 was used for dose measurements. The dose was measured using a human phantom (Alderson Research Laboratories, USA) in the CT scan with a 3rd generation CT scanner of Somatom Plus (Siemens, Germany) for bone imaging. CTDI for this CT scanner were 9.2mGy/ 100mA at the center in an acrylcic resin phantom with diameter of 16cm and 8.5mGy/100mA at 1cm depth from the phantom surface. X-ray tube voltage of 120kV and tube current of 85mA was used. Slice thickness was varied from 1 to 3mm and table speed per rotation was also varied from 1 to 5mm per rotation. X-Omat-V (Eastman Kodak, USA) films and TLD (Thermo-Luminescent-Dosimetry) dosimeters of the type of MSO-S (Kyokko, Japan) were used in the dosimetry.
Result: Patient radiation dose reduced with increasing the pitch of SPIRAL scan. Measured dose was uniformly distributed and well corresponded to the dose calculated using CTDI. However, measured doses on scanning with 1mm slice thickness were always higher than those with 2 to 5mm slice thickness. The lowest radiation dose was obtained with scanning with 2mm slice thickness and table speed of 4mm per rotation which give the dose of about 4mGy per one CT examination in the imaged tissues. The highest dose per one CT examination was measured in “dental CT” for the mandibular region with 1mm slice thickness and table speed of 1mm per rotation which gave 12mGy by film dosimetry and 9 mGy by TLD dosimetry.
Conclusion: SPIRAL scan with pitch more than 1 was effective for reduction of patient radiation dose without reducing the image quality. CTDI was also useful to estimate the dose except scans with 1mm slice thickness.
View full abstract