A conversational data processing system has been developed for the analysis of spark source mass spectra on a photoplate.
This is the hierachy system with the IBM S/370and the S/7. The S/7is used for the data acquisition and the first data processing.
The calibration curve for the photoplate is based on the Hull empirical relationship, and the determination of ion intensity is calculated by the integration method of peak area intensity.
The time required for the quantitative analysis by this system is about one tenth of that by the conventional method.