The detector of the MSO-S type of the thermoluminescnce dose meter depends on energy. Several types of filters to the sensor are added in order to develop a method for facilitating the evalution of radiation dose in cases of multiple levels of radiation energy. Four types of filters are made for the experiment. These filters were assembled to a single detector, which is called the MSO-Y type detector.The sensitivity of the dose meter including the detector for radiation energy are measured. On the basis of the sensitivity, we investigated how to measure the effective energy, how to calculate the calibration constant of energy, and how to evaluate the radiation dose. As the result, it was found that this radiation measurment method is effective in the range of X-ray energy comparable to or less than the radiation energy for the radiography.