A differential optical absorption spectroscopy (DOAS) system for measurement of atmospheric NO
2 was developed. The system uses a battery-operated, high luminance LED and a fiber-coupled spectrometer, and is portable. Laboratory experiments using a gas cell of length 0.22 m with varying NO
2 concentrations were performed to evaluate the sensitivity of the DOAS system. The DOAS measurement results are in agreement with NO
2 concentrations obtained simultaneously by a FT-IR (Fourier Transform Infrared) system for NO
2 concentrations down to 20 ppm. Experiments with an optical path length of 93 m were also performed, and NO
2 concentrations down to 0.20 ppm were measured. Since measurement of atmospheric NO
2, which is in the order of several tens of ppb, requires optical path lengths of several hundred m, system improvements to improve the signal detection are necessary.