Japanese Journal of Radiological Technology
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
Volume 76, Issue 4
Displaying 1-14 of 14 articles from this issue
Opening Article
Originals
  • Kenji Matsumoto, Masakazu Otsuka, Mikoto Tamura, Hajime Monzen, Masahi ...
    2020 Volume 76 Issue 4 Pages 339-345
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    Purpose: Novel linac improvements in speed of gantry, collimator, leaf and dose rate may increase the time-efficiency of volumetric modulated arc therapy (VMAT) delivery, however remains to be investigated. In this study, a fast-rotating O-ring linac (Halcyon) with fast moving leaves is compared with a general linac (TrueBeam: TB) in terms of plan quality for VMAT of C-shape, prostate, multi target and, head and neck (H&N) cases from AAPM TG-119. Materials and Methods: For the four test cases, VMAT planning was performed using single to four-arc VMAT on a Halcyon and using single to three-arc VMAT on a TrueBeam. Same conditions for optimization were used in each test case. Target coverage metrics and organ at risks (OAR) dose were compared. Monitor unit (MU) and irradiation time in each plan were also compared. Results: In all cases, single-arc plans of Halcyon were inferior to TB plans on dose objectives. Conformity index (CI) to outer target of C-shape case was better for Halcyon (1-arc: 1.242, 2-arc: 1.202, 3-arc: 1.198, 4-arc: 1.181) than for TB (1-arc: 1.247, 2-arc: 1.211, 3-arc: 1.211) except to single arc. D5 (Gy) of core for C-shape case was better for halcyon (1-arc: 23.29, 2-arc: 21.01, 3-arc: 20.64, 4-arc: 20.47) than for TB (1-arc: 24.04, 2-arc: 22.94, 3-arc: 23.04). Calculated MU was smaller for Halcyon than for TB. In addition, Halcyon is more faster than TB because mechanical movements were improved. Conclusion: For VMAT plan in each case, Halcyon as well or better at the plan quality of two or three arcs on TB while reducing the delivery time.

    Download PDF (1157K)
  • Yusei Nishihara, Masanao Kobayashi, Hiroki Saito, Tomonobu Haba, Yasuk ...
    2020 Volume 76 Issue 4 Pages 346-355
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    Presently, the scanning start angle of the X-ray tube of X-ray computed tomography (CT) scanners cannot be controlled. As a result, there is room for reducing patient dose because the peaks of the dose distributions may overlap during multiphasic CT imaging. This study investigated methods of dose reduction by performing a Monte Carlo simulation of the X-ray tube scanning start angle and locally absorbed dose in multiphasic CT imaging. In the Monte Carlo simulation, the largest decrease in the absorbed dose was seen, when the scanning start angle between the phases was±180°. Even though with present X-ray CT scanners, the scanning start angle cannot be controlled, it is possible to decrease the absorbed dose by taking the orbital synchronized scanning and scanning range into consideration. In future we hope that, we will be able to easily reduce the dose by controlling the scanning start angle.

    Download PDF (2845K)
  • Tatsuya Kamima, Yu Murakami, Masato Arima, Yosuke Sato, Minoru Yoshiok ...
    2020 Volume 76 Issue 4 Pages 356-365
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    Purpose: The aperture shape controller (ASC) decreases the complexity of the multi-leaf collimator (MLC) aperture for volumetric modulated arc therapy (VMAT). The purpose of this study was to evaluate the effect of ASC settings on the VMAT plan quality. Method: First, VMAT plans were created (ASC=off) for three test patterns of The American Association of Physicists in Medicine (AAPM) Task Group 119 (TG-119) and 20 cases of nasopharyngeal cancer. Next, for these VMAT plans, only the ASC settings were changed from very low (complexity reduction: low) to very high (complexity reduction: high) in five steps, and VMAT plans were created respectively. To evaluate the created VMAT plans per each ASC settings, we analyzed the modulation complexity score (MCSV) and dosimetric parameters for the planning target volume (PTV) and organ at risk (OAR). Result: In three test patterns, there were no major dosimetric differences between the VMAT plans. In nasopharyngeal cancer, the mean MCSV were 0.413, 0.325, 0.320, 0.307, 0.303, and 0.272 for very high, high, moderate, low, very low, off settings, respectively. Therefore, the most complex MLC aperture was off, followed by very low, low, moderate, high, and very high. In terms of dosimetric parameters, the VMAT plans created using the very high setting showed an increase of D2% in the PTV and worse OAR sparing than that using other ASC settings. On the other hand, the dosimetric results for the very low to moderate setting obtained similar results to those for the off setting, respectively. Conclusion: The ASC was able to decrease the complexity of the MLC aperture according to the setting level. From very low to moderate settings, a plan equivalent to the off setting could be created in terms of dose parameters.

    Download PDF (3293K)
Note
  • Nobuo Kitera, Kosuke Matsubara, Chikako Fujioka, Kazushi Yokomachi, Ei ...
    2020 Volume 76 Issue 4 Pages 366-374
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    Purpose: Organ-based tube current modulation (OB-TCM) techniques, which are provided by three vendors, reduces the radiation dose to the lens of the eyes by decreasing the tube current, when the X-ray tube passes over the anterior surface of critical organs. However, the characteristics of dose modulation of these techniques are different. The purpose of this study was to understand the performance characteristics of OB-TCM technique of each computed tomography (CT) vendor at head CT. Methods: We used three CT scanners (SOMATOM Definition Flash; Siemens Healthcare, Revolution CT; GE Healthcare, and Aquilion ONE Genesis Edition; Canon Medical Systems). We measured the radiation dose to the lens surface as evaluation of radiation dose reduction and measured the image noise as index of image quality. We measured the radiation dose rate in the air for analysis of the characteristics of dose modulation in each OB-TCM. Results: When applying OB-TCM, the radiation doses for the lens surface were decreased by 28%, 22%, and 25% for Siemens, GE, and Canon CT scanners, respectively, and the image noise level was increased by 5.6%, 8.5%, and 15.1% for Siemens, GE, and Canon CT scanners, respectively. The characteristics of dose modulation in each OB-TCM were also confirmed by measured the radiation dose rate. Conclusion: We confirmed that each OB-TCM has different influence on image quality and radiation doses for lens surface, due to the different characteristics of dose modulation for each CT vendor.

    Download PDF (1593K)
Clinical Technologies
  • Haruhide Kunimoto, Hideharu Miura, Masanori Miyazawa, Masahiro Hayata, ...
    2020 Volume 76 Issue 4 Pages 375-384
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    In order to correct the lateral effect caused by the light source of the flatbed scanner in the Gafchromic film EBT3, the usefulness of the correction method using the average value of the correction coefficient considering the scan directions were evaluated. EBT3 was scanned from four directions to measure the optical density (OD) of the red, blue, and, red/blue components and the correction coefficient were calculated. For the correction coefficients, average values were calculated for the purpose of use, when the scan directions could not be aligned (average lateral effect correction). Correction accuracy was verified with the pass rate of gamma analysis (3 mm/3%, threshold 30%) of the dose distribution using the EBT3 film irradiated with the step pattern. OD of the red, blue, and, red/blue components in the scanning vertical direction tended to be higher in the center than in the peripheral portion. The pass rate of the step pattern was the red component’s before correction, from 26.9 to 45.1% (before correction), from 84.1 to 96.7% (after correction), the red/blue component, from 37.6 to 48.4% (before correction) and from 84.4 to 96.7% (after correction). When using the correction coefficient using the average value, the pass rate was 89.8% for the red component and 94.7% for the red/blue component. The lateral effect correction improves the accuracy of the dose distribution verification, and the correction coefficient using the average value is useful when the scanning direction is different from that at the time of obtaining the dose concentration curve.

    Download PDF (5302K)
  • Yuuki Yoshimura, Daisuke Suzuki, Kanae Miyahara
    2020 Volume 76 Issue 4 Pages 385-393
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    The upper abdomen was imaged with diffusion weighted images for free breathing and respiratory suppression using single shot-echo planar imaging (SS-EPI) and readout segmented-EPI (RS-EPI). We examined the usefulness of respiratory suppression imaging for the subject of healthy volunteers. Motion artifacts, apparent diffusion coefficient (ADC) values, and organs movement distances were evaluated. As a result, motion artifacts and organs movement distances were reduced in respiratory suppression than free breathing. The ADC values did not change. Respiratory suppression was simple and useful. In addition, it was found that RS-EPI imaging could be used for imaging the upper abdomen in the same way as SS-EPI by respiratory suppression.

    Download PDF (2294K)
Data
  • Hiroo Segawa, Akihiro Miki, Tatsuya Yamasaki, Yasuhiro Ide, Yuichirou ...
    2020 Volume 76 Issue 4 Pages 394-403
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    In anatomical charts in conventional books, the pathways of nerve fibers are drawn in illustrations. Conversely, with diffusion tensor tractography (DTT), we can visually understand the trajectory of nerve fibers through color. We created a stereo color anatomical chart of the nerve fibers that can be used for magnetic resonance (MR) examination to diagnose the pathway of nerve fibers and that can be used to explain the results of MR examination to visually understand how nerve fiber information is transmitted from the frontal lobe, parietal lobe, occipital lobe, temporal lobe, cerebellar lobe, and cerebral cortex.

    Download PDF (12970K)
  • Kazuya Shinoda, Koichi Shida, Shogo Hatanaka
    2020 Volume 76 Issue 4 Pages 404-409
    Published: 2020
    Released on J-STAGE: April 20, 2020
    JOURNAL FREE ACCESS

    Multi-leaf collimator (MLC) parameters, which are registered with radiation treatment planning systems, are very important for intensity modulated radiation therapy (IMRT). In this study, we investigated MLC parameters of respective institutions for efficient commissioning of IMRT. Data of linac models, MLC types, nominal energy, irradiation technique, calculation algorithm, dosimetric leaf gap (DLG) values, and MLC transmission values were collected from each institution in which Varian linac and Eclipse were owned, and analyzed. The numbers of responses from institutions to questionnaire were 15, and the total number of linac was 22. In most institutions, volumetric modulated arc therapy (VMAT) technique was used, and the most used nominal energy was 10 MV. The higher nominal energy was, the higher values of MLC parameters (DLG and MLC transmission) were. In addition, values of MLC parameters of flattening filter free (FFF) beams were smaller than those of flattening filter (FF) beams, even when nominal energy was same. Values of DLG of VMAT tended to be greater than those of multi-field IMRT. These results are expected to be useful for institutions, in which IMRT is implemented.

    Download PDF (742K)
Educational Lecture-Roadmap for Safer Installation of Medical Accelerators-
Educational Lecture-Utilization of ImageJ in Radiological Technology-
Educational Lecture-How to Proceed Research in Radiological Technology: From Planning a Research Design to Submit a Research Article-
Basic Lecture-Quantitative Evaluation and Associated Uncertainties in Radiological Technology-
JIRA Topics
feedback
Top