In modern radiography, mA-sac meters are commonly used to measure the amount of X-ray exposure which is the product of the tube current and the time of application. It has been thought almost impossible to eliminate the tau-characteristics (the strong dependency of indica-tion upon the time) of common mA-sec meters of ballistic type as they are used in wide range of time, overstepping the axiom of ordinary ballistic galvanometers. Instantaneous exposure as required in X-ray cinematography and stereo-photography made the realization of accurate integrators imperative.
The author describes the developement of a theory and technique leading to the realization of deviation-free integrator (Fig. 6) and, at the same time, discussions on integration network which exhibits a high grade of accuracy of the third order of precision (Fig. s 13 and 15) and has an attractive behaviour of the indication memory for a long time. To calibrate the meter in absolute units, the author constructed a simple cybernetics system (Fig. 9), nervous system of which displays the power of quick and correct calculation and of good memory. High accuracy in radiographic technique is thus expected.